Service Hotline
400 8392 001
1. Full Chinese LCD display, friendly man-machine interface, simple keyboard operation, easy to learn and use.
2. PID mode control, high temperature control accuracy, can effectively protect silicon carbon tubes, prolong service life.
3. Setting furnace temperature and correcting can be changed directly by keyboard input.
4. Intelligent control of electrolytic switch by software will not cause over-electrolysis due to misoperation.
5. Furnace temperature, flow, indicating voltage, process time and sulfur content can be checked at any time. The test process is clear at a glance.
6. Automatically judge the titration end point, shorten the experiment time and improve the work efficiency.
7. Input of water value, the dry base sulfur and water value can be calculated and printed.
8. Store the latest 30 historical records and print them at any time.
9. The working temperature of the system can be set between 600 ~1150 ~C by the software (1050 ~800 ~C for coal and 800 ~C for petroleum).
10. The standard RS232 serial interface is used between the microcomputer and the single chip computer system, which has the characteristics of simple connection, reliable transmission, long transmission distance and low failure rate.
11. The single-chip computer system is responsible for controlling the temperature of the high temperature furnace and collecting the measured data. The data are transmitted to the main control computer through the serial line. The data are processed by the computer, and the overall performance of the system is greatly improved with the cooperation of the main program of the computer.
12. The system places the high temperature furnace, magnetic mixer, air purification device, electrolytic cell, automatic sample feeding mechanism and single-chip computer control circuit in the same case reasonably, which makes the whole system beautiful, small, easy to place and connect, simple operation and daily maintenance of the system.
The micro-computer fast integrated sulfur meter consists of air purification device, controller, combustion furnace, electrolytic cell and agitator. It is mainly used to determine the total sulfur content in coal, iron and steel and various minerals. It is the preferred and necessary instrument for laboratories of departments such as coal, power, chemical industry, building materials, metallurgy, geological exploration, commodity inspection and environmental protection detection.
Adoption of standards
GB/T 214-2007 "Determination of Total Sulfur in Coal"
GB/T 483-2007 "General Provisions for Coal Analysis Experiment Method"
GB/T 387-1990 Tube Furnace Method for the Determination of Sulfur Content in Dark Petroleum Products
Measurement range: 0.1%-20%
Analysis time: about 5 minutes
Temperature Control Accuracy: +5 Temperature
Test temperature: 1150 C
Temperature Control Accuracy: +5 Temperature
Power: <4KW
Power supply: AC (220 + 22) V 50Hz