Service Hotline
400 8392 001
1, the calorimeter system can automatically standard the heat capacity of the calorimeter system.
2, the use of imported stirring motor, stirring uniform speed and stability, reliable performance, strong anti-interference ability. Realize automatic water filling, automatic water temperature adjustment, automatic quantitative water, automatic stirring, automatic ignition, reduce human error. This machine can be one machine double control, do not interfere with each other.
3, the calorimeter device adopts electric stirring of blade in the inner cylinder, and submersible electric stirring in the outer cylinder, so that the mixing is more uniform and more convenient.
4. The case shell has a water level indicator line, which can observe whether the inner cylinder is short of water at any time.
5, outside the chassis can be observed at any time ignition voltage, ignition current.
6, the experiment automatic cooling correction and the use of double barrel water cycle, higher thermal stability, higher test accuracy, the inner barrel using advanced foam heat insulation technology, the experiment process is not affected by external temperature. It can continuously analyze samples for a long time and has a very high cost performance.
7, controlled by microcomputer, only need to input sulfur, moisture, hydrogen and other data, you can convert and print out the shell heat, high, low heat and other data. Automatic calculation and automatic conversion of high and low heat, can print out a regular coal quality test report, automatic storage of data up to 999.
8. The system can automatically complete the calibration of system energy equivalent and the determination of material calorific value. The calibration of system energy equivalent adopts multi-dimensional selection method and is real-time and flexible. The measurement process and data processing of material calorific value are automatically completed by microcomputer
9, support ignition wire and cotton two ignition methods
10, the measurement process using digital prompts and images, image intuitive.
11, calorimeter can automatically display ignition voltage and ignition current, can detect the heat of lower or higher substances, with automatic diagnostic function.
High-precision automatic calorimeter is mainly suitable for determining the calorific value of solid or liquid combustible materials such as coal, coal, paper, petrochemical, cement, agriculture and animal husbandry, medicine, scientific research, teaching and other industries or departments.
Meet the standard:
GB/T483-2007 "General Regulations on Coal Analysis Experimental Methods"
GB/T 213-2008 "Coal calorific value determination method"
GB/T 384-1981 "Determination of calorific value of Petroleum products"
GB/T 30727-2014 "Determination of calorific value of solid biomass fuel"
GB/ T14402-2007 "Combustion properties of building materials and products - Determination of combustion calorific value"
GB/T 30991-2014 "General Technical Conditions for Intelligent oxygen bomb calorimeter"
JC/T 1005-2006 "Cement black raw material calorific value determination Method"
JJG 672-2001 "Verification Regulations of Oxygen bomb Calorimeter"
ASTM D5865-2013 - Determination of Calorific Value of Coal and Coke
ASTM D5865-2007 Standard Test Method for Total Calorific Value of Coal and Coke
ISO 1928-2009 Solid mineral fuels - Determination of total value and calculation of net calorific value by bomb calorimeter
CEN/TS 14918 "Determination of Calorific value of solid biofuels"
BS EN 15400-2011 "Solid Recovered fuel - Heat Value Test"
IS:1350-1970 "Method of determination of Coal and Coke"
1. Use environment: 0-65℃
2. Precision of heat capacity: ≤0.1%
3. Resolution: 0.0001℃
4. Measurement accuracy: meet the requirements of GB/T 213-2008
5. Test time: about 10min
6. External water cylinder capacity: 75L(double barrel)
7. Inner water cylinder capacity: about 2.3L
8. Ignition voltage: 24V
9. Ignition time: 3 minutes for ignition
10. Measurement error: better than the national standard (<10 calories)
11. Host power consumption: ≤ 40W
12. Power supply: 220 V(-15%-10%)50Hz
13. Host size: 830mm×560mm×470mm