1. Fast: the infrared absorption method is used, the measurement speed is fast, and the time to test a sample is only about 400s; It can be loaded into 36 samples at one time for continuous measurement, and the samples can be inserted in the measurement, which is suitable for the analysis of a large number of samples.
2. Accuracy: high temperature combustion, oxygen combustion, to ensure the complete combustion of the sample and the full release of carbon, hydrogen, nitrogen elements, take effective measures to overcome the adverse effects of environmental factors, good repeatability, high accuracy.
3. Convenient: high degree of automation, by computer control to automatically complete the sample delivery, measurement, calculation results, no need for manual intervention; Can display, recalculate, save, query all results; With statistical function, it can easily obtain important data such as mean value, range, standard deviation and relative standard deviation of the result; When the computer is connected with the electronic balance, the weight of the sample can be directly transmitted, without manual input. The test data can be transmitted to the Internet through RS232 interface. No need to prepare chemical reagents and electrolytes.
4. Reliable: selection of high-quality components and materials, take a variety of anti-interference measures, with overtemperature protection, overload protection and leakage protection, the whole machine performance is stable, safe and reliable.
The hydrogen nitrogen analyzer is suitable for determining the content of carbon, hydrogen and nitrogen in coal and coke. The hydrogen measured by this instrument is total hydrogen in the substance, including organic hydrogen, hydrogen in water and hydrogen in mineral crystal water. Carbon is the total carbon in the material, including organic carbon and inorganic carbon, etc. Nitrogen is the total nitrogen in a substance. It is suitable for power plants, coal mines, metallurgy, chemical industry, commodity inspection, scientific research, teaching and other industries.
1. Analysis method: infrared spectroscopy for carbon and hydrogen, thermal conductivity method for nitrogen.
2. Analysis range: hydrogen content 0.05% ~ 50%, carbon content 0.05% ~ 100%, nitrogen content 0.008% ~ 50%; The recommended sample weight is 75mg and the particle size is less than 0.2mm.
3. Analysis time: about 400s.
4. Repeatability: Had is not more than 0.15%, Cad is not more than 0.5%, Nad is not more than 0.08%.
5.Reproducibility: Hd is not more than 0.25%, Cd is not more than 1%, Nd is not more than 0.15%.
6. Furnace temperature: the main combustion furnace 950℃, (900 ~ 1100) ℃ can be set, the error does not exceed the set value of ±1%. Secondary combustion furnace 850℃, (800 ~ 1000) ℃ can be set, the error does not exceed the set value of ±1%. Reduction furnace 600℃, (550 ~ 650) ℃ can be set, the error does not exceed the set value of ±1%.
7. Alarm temperature: When the main combustion furnace, secondary combustion furnace, reduction furnace and gas storage tank overtemperature, the heating power supply is automatically cut off.
8. Constant greenhouse temperature: 50℃, (45 ~ 52) ℃ can be set.
9. Gas tank temperature: 50℃, (45 ~ 52) ℃ can be set.
10. Oxygen: purity 99.5%, pressure (0.25±0.01) MPa, the total pressure of the air source is not less than 1MPa, can not use electrolytic oxygen, oxygen flow rate is divided into 0.5L/min, 1.5L/min, 3L/min three grades.
11. Power gas: nitrogen or compressed air (need to remove oil and water), pressure (0.25±0.01) Mpa.
12. Carrier gas: Helium, purity is higher than 99.996%, pressure (0.25±0.01) Mpa
13. Furnace reagent: special reagent with particle size of 2mm ~ 5mm.
14. Electronic balance: sensitivity 0.0001g
15. Computer: Lenovo Business Computer, windowsXP operating system, Pentium 733MHZ, 256MB RAM, USB interface.
16. Weight: about 85kg
17. Main machine dimensions: 775mm(L)×575mm (W)× 780mm (H)